Manufacturer record

Supplied by M/s. Kerala Medical services Corporation Limited: regulatory record

Last checked 3 November 2026. Source: CDSCO monthly NSQ alerts and state drug-testing laboratories

2 batches manufactured by Supplied by M/s. Kerala Medical services Corporation Limited (Kerala) were declared Not of Standard Quality in April 2024. Records are batch-specific findings by government laboratories as published in CDSCO's monthly alerts; where a company has stated that a flagged batch was counterfeit, that statement is noted when reported.

Most recent flagged batch: D 2259/23- 24, Salbutamol Nebuliser Solution BP (exp May 2025). SDTL Thiruvananthapuram (state lab) found it not of standard quality: It fails the test Acidity pH as per BP. Listed in the April 2024 alert.

Records by type

2 records linked to Supplied by M/s. Kerala Medical services Corporation Limited in the datasets we hold. A zero means nothing is on record in those datasets, not that nothing ever happened. What the labels mean.

0Banned or prohibited
0Recalled or suspended
2Failed a lab test
0Regulatory notice
Primary government sourceTier 1: government primary source

CDSCO drug alerts: Not of Standard Quality and spurious drugs

CDSCO · CDSCO monthly NSQ alerts, Apr 2024 to Apr 2024

View original document

Scope. A company appearing here is not a finding that its products are unsafe. Regulatory actions can apply to one batch, one formulation, one period or one state. Always check the linked government document for the current position, and tell us if something is out of date. How we label records.

2
NSQ batches since 2024
2
Products affected
Sterility
Most common failure
Apr 2024
Latest alert

Flagged batches per year

2024 · 2

All flagged batches

Product Batch Reason declared NSQ Tested by Alert
Salbutamol Nebuliser Solution BP Batch D 2259/23- 24Mfg Jun 2023 · Exp May 2025 It fails the test Acidity pH as per BPpH SDTL ThiruvananthapuramState lab Apr 2024
Bupivacaine Hydrochloride in Dextrose Inj. USP 0.50% 4 ml Amp Batch D2258/23- 24Mfg Sep 2023 · Exp Aug 2025 It fails the test Sterility as per USPSterility SDTL ThiruvananthapuramState lab Apr 2024

Right of reply. Manufacturers may submit a response or a laboratory re-test result; it is published unedited alongside this record. Submit a response.